Fluke 190B/C MedicalUser Manual2-2
3Display the
SCOPEkey labels
4Open the Reading 1 menu
5Select
on A, then accept
6Select
mAs, then accept.
This opens the Current Probe menu.
7Select for example
1 V/A, then accept.
After closing the information
banner the screen shows(see figure 2-2) :
• Top left reading,
mA (or AA, etc.):
the difference between the current amplitude measured on the left cursor and measured on the
right cursor Top right reading,
mAs (or As, As, etc.):
the integrated current over the time between the cursors.
Figure 2-2. mAs measurement screenNote: the B versions F key label does not show RMS asthese versions do not provide RMS measurementson the trace section between the cursors.8Highlight the left or right cursor.
9Move the highlighted cursor to the desired position on the waveform.
Shop for Fluke products online at:
1.888.610.7664www.
MyFlukeStore.com
Using the Medical FunctionsmAs Measurements2 From step 2, you can also proceed as follows to do mAs measurements on input A:
3In
Scope mode display the INPUTA key labels.
4Open the
Probe on A menu.
5Select
Probe Type Current,
then
accept.
6Select for example
1 V/A, then accept.
7Display the cursor key labels (Band C-versions are different).