Ansi c63. 19 -2a -2007 Revision of


Annex C (normative) Equipment and setup calibration



Download 1 Mb.
Page9/25
Date10.08.2017
Size1 Mb.
#31128
1   ...   5   6   7   8   9   10   11   12   ...   25

Annex C
(normative)
Equipment and setup calibration

C.1Test enclosures

C.1.1RF test enclosures


When tests require semi-anechoic enclosures, the quality of the enclosures shall be measured. This can be done using at least two common ways as follows:


  1. Measuring and meeting the “volumetric” normalized site attenuation using the procedures described in ANSI C63.4

  2. Calibrating the radiated immunity field using the procedures in IEC 61000-4-3

The previously mentioned general requirements also satisfy the requirement that the undesired reflections from those enclosures, which are not perfectly anechoic, are suppressed by at least 20 dB. In addition, such anechoic enclosures shall be tested to ensure that any RF leakage from the RF ambient outside the enclosure is at least 10 dB down from the required measurement level.


The calibration requirements for wideband transverse electromagnetic (WB TEM) devices are based on the direct measurement of the field produced in the test zone. A secondary procedure is performed by calculation, using the power into the device and the physical dimensions of the center conductor (septum) separation from the referenced ground plane. See [B2], [B9], [B10], ANSI C63.4, and FCC 47 CFR 20.19.

C.2Audio input source


An acoustic signal shall be supplied to the hearing aid under test during application of interfering RF test fields, so that hearing aid automatic gain control circuits are in a realistic state of operation, and so that interfering effects on the intended audio signal may be detected. This frequency shall have sufficient separation from 1000 Hz (the RF modulation frequency used in testing) so that as the intended signal it can be independently observed at the hearing aid output by means of tuned filters or an audio signal analyzer or wave analyzer.
A frequency of 1300 Hz shall be used provided that the hearing aid frequency response between 1000 Hz and 1300 Hz is smooth (< 0.5 dB ripple) and changes less than a total of 2 dB over this range. The pre-test data shall be examined to verify this fact. If necessary one or both of the two frequencies involved may be shifted to a region where the response is smoother and the slope is less than 6 dB per octave in the frequency interval. The frequency interval should be at least 1/3 octave and should lie within the bounds of 600 Hz to 3000 Hz.

C.3Calibration of RF E-field and H-field probes


The purpose of the calibration for probe modulation response factor is to align the probe readings with the quantity of the RF signal most closely correlated with the intensity of interference to hearing aids.
Probes used for measuring near-field emissions and calibrating immunity field levels shall be calibrated using the guidelines contained in IEEE Std 1309-2005. The field pattern shall be isotropic to a tolerance of
± 20%. If probes with coaxial cables are used, the influence of cables on the field shall be accounted for in the calibration. The H-field probe may have one, two, or three loops. In the case where these probes have less than three mutually orthogonal dipoles or loops, they shall be capable of rotation about their geometric centers to allow calibration and measurement in the three orthogonal axes.
IEEE Std 1309-2005 provides for three calibration methods and several grades of calibration. The two methods that are most appropriate for ANSI C63.19 probe calibrations are Method A, using a transfer standard probe, or Method B, using a standard gain horn antenna in an anechoic chamber. The most appropriate grade of calibration is: FD, F2 (fL, fM, fH),1 A1, I0 for single-axis probes or I1 for “isotropic” probes, R0, T0, and M0. The grade designations are found in Annex A of IEEE Std 1309-2005.
The “maximum interception alignment” as defined and specified in 4.2.2.1 of IEEE Std 1309-2005 shall
be used for calibration. The calibration field generation shall be via a pyramidal horn antenna in an anechoic chamber, either as a standard gain antenna, Method B (Table 2 of IEEE Std 1309-2005), or
as a reference field generator using a similar probe as a transfer standard, Method A (Table 2 of
IEEE Std 1309-2005). Anisotropy of “isotropic” probes shall be measured in accordance with 7.1.3 and Equation (2) of IEEE Std 1309-2005.
The best-case expanded uncertainty, U, for probe calibration is U ≈ ± 1.1 dB for Method A and U ≈ ±
1.0 dB for Method B. Out of the allowable expanded uncertainty of ± 2 dB (see 4.1.2), these calibration uncertainties leave approximately ± 0.84 dB to ± 0.86 dB in terms of combined standard uncertainty, uC, for other contributors. That is: U = ± 2 dB, thus uC = ± 2 ÷ 2 = ± 1 dB; uS2 = uC2ucal2; ucal = Ucal ÷ 2 =
± 1.1 ÷ 2 = ± 0.55 dB (the value, Ucal = ± 1.1 dB, is from Method A); and, uS = √(12 – 0.552) = ± 0.84 dB. (See Annex E for further information on estimation of uncertainty.)

C.3.1RF field probe modulation response


In addition, for probes with a response to variations in the RF field of < 20 kHz, a calibration shall be made of the modulation response of the probe and its instrumentation chain. This calibration shall be performed with the field probe attached to the instrumentation that is to be used with it during the measurement. The response of the probe system to a CW field at the frequency(s) of interest is compared to its response to a modulated signal with equal amplitude. The field level of the test signals shall be more than 10 dB above the ambient level and the noise floor of the instrumentation being used. The ratio of the CW reading to that taken with a modulated field shall be applied to the readings taken of modulated fields of the specified type. This may be done using the following procedure:


  1. Fix the probe in a set location relative to a field generating device, such as a reference dipole antenna or WB TEM, as illustrated in Figure C.1.

  2. Illuminate the probe with a CW signal at the intended measurement frequency.

  3. Record the reading of the probe measurement system of the CW signal.

  4. Record the power level of the CW signal being used to drive the field generating device.

  5. Substitute a signal using the same modulation as that used by the intended WD for the CW signal.

  6. Set the amplitude during transmission of the modulated signal to equal the amplitude of the CW signal.

  7. Record the modulated signal reading from the probe measurement system.

  8. The ratio, in linear units, of the CW to modulated signal reading is the modulation factor.



Figure C.6—Dipole calibration procedure

An alternative procedure is as follows:




  1. Fix the field probe in a set location relative to a field generating device, such as the reference dipole antenna, as illustrated in Figure C.1.

  2. Illuminate the probe using the wireless device connected to the reference dipole with a test signal at the intended measurement frequency, Ensure there is sufficient field coupling between the probe and the antenna so the resulting reading is greater than 10 dB above the probe system noise floor but within the systems operating range.

  3. Record the amplitude applied to the antenna during transmission and the field strength measured by the E-field probe located near the tip of the dipole antenna.2

  4. Replace the wireless device with an RF signal generator producing an unmodulated CW signal and set to the wireless device operating frequency.

  5. Set the amplitude of the unmodulated signal to equal that recorded from the wireless device.

  6. Record the reading of the probe measurement system of the unmodulated signal.

  7. The ratio, in linear units, of the probe reading in Step 6) to the reading in Step 3) is the E-field modulation factor.

  8. Repeat the previous steps using the H-field probe, except locate the probe at the center of the dipole.

The modulation factors obtained by one of these methods shall be applied to readings taken of the actual WD, in order to obtain an accurate reading.
In the procedures above, the amplitude of the signals shall be set so that the resulting reading is greater than 10 dB above the probe system noise floor but within the systems operating range.
When performing this procedure the operator must ensure that discontinuous transmission (DTX) is disabled, or some means of preventing the WD from switching into the DTX mode must be employed. Depending upon the measurement method utilized, failure to disable DTX can result in a substantial measurement error.


Download 1 Mb.

Share with your friends:
1   ...   5   6   7   8   9   10   11   12   ...   25




The database is protected by copyright ©ininet.org 2024
send message

    Main page