Automatic analytical methods for environmental monitoring and control



Download 4.48 Mb.
Page2/19
Date31.03.2018
Size4.48 Mb.
#44836
1   2   3   4   5   6   7   8   9   ...   19

Figure III.2.1. Diagram of CFA system. (a) continuous mixing method; (b) stopped-flow continuous mixing method – reagent A is continuously aspirated and mixed with the alternating flow sample S and carrier C; (c) continuous-flow titration – P1 – pump of constant speed, P2 – pump of variable speed, controlled by computer. P – pump; RC – reactor (reaction coil); Ddetector; KL - kinematically controlled probe; C – carrier; S – sample; A, B – reagents; W – waste; ttr – titration time.

  1. Continuous-flow titrations – the sample is continuously inserted into the system either by keeping its speed constant and changing that of the titrant (Figure III.2.1.c), or both sample and titrant can be kept at a continuous speed and measurements can be made as a function of the analytical signal obtained.


Applications of CFA systems.

The continuous monitoring of cyanide anion, which is a highly toxic ion, has been carried out by using a CFA system (Figure III.2.2) and a classical analytical method with spectrophotometric detection (barbituric acid and chloramines T). As Figure III.2.2 illustrates, the stream of chloramines T is mixed with the sample then merged with a pyridine/barbituric acid stream after the corresponding reaction coil. A second reaction coil allows the colored reaction product to form and be measured at 578 nm. This method, also realized by normal FIA and reverse FIA, permits a comparison between these two modes and CFA.






Figure III.2.2. Diagram of CFA system used for cyanide spectrophotometric determination. P – peristaltic pump; RC – reaction coil; D – detector; W – waste.

III.2.1.2. Segmented Flow Analysis
The segmented flow analysis (SFA) was the earliest contribution in the field of automated methods development. It originated from the scientific paper of Dr. L Skeggs (University of California, USA) published in 1957 and has found widespread use in almost every facet of analytical chemistry. Skeggs’s studies were materialized in the design of the first continuous dynamic measuring system with sequential introduction of samples and the use of a flow-cell. Sample carry-over was prevented by segmentation with air bubbles introduced between successively aspirated samples.

Several international corporations, such as Technicon, Skalar, Burkard, etc. have contributed to the development and commercialization of analyzers and assemblies based on Skeggs’s idea. For many years, these were the only alternative available for the automation of high-throughput control laboratories.


General aspects and instrumentation

SFA is characterized by the use of one or several liquid streams where the samples are introduced by sequential aspiration and separated by means of air bubbles aimed at avoiding the carry-over. Therefore, the final liquid stream is segmented into small discrete liquid slugs by bubbles of air or other gas that entirely fills the stream tubing bore.

The sample and reagents are mixed by passing through glass coils and through a temperature controlled heating coil if heat is required to speed the development of the reaction product before detection. In the initial work, it was found that in some analyses the high molecular weight components contained in samples were interfering with the chemical reactions. This problem was ingeniously overcome by the use of a cellophane dialysis membrane to remove them. Today, the technique has become one of the most reliable and widely used methods for automatic chemical analysis in routine and research analytical laboratories. This technique has the advantage of for measuring large batches of samples for up to 16 analytes simultaneously at speeds of up to 120 samples.

A SFA system is presented in Figure III.2.3 and it comprises a series o modules each performing a specific function, e.g. sampling, propelling device, sample transport, heating, separation unit, detector equipped with a flow-cell, data recording and analysis module, all of them coupled on-line to one another.






Figure III.2.3. Scheme of a SFA system.
Sampling device – which consists of a sample turntable and moving articulated aspiration probe.

Propelling device – aimed to provide the sample and reagent and air streams. These are generally multi-channel peristaltic pumps (Figure III.2.13) but may also be used piston pumps and the pressure exerted by a gas or gravitational force. The flow-rate of the streams can be adjusted and maintained as constant as possible by using flexible plastic tubes that withstand the mechanical pressure to which they are subjected.

Reaction-mixing coils – sample and reagents merge in the appropriate stages through “T”-connectors and then pumped through the glass, PTFE or polyethylene tubing where the mixing of reactants and the analytical reaction takes place. The tubing is coiled and horizontally mounted. It provides the reactants mixing by repetitive inversions of the liquid phase and its length governs the time over which the reacting mixture ‘resides’ in it and hence the sampling frequency.

Heating device, continuous separation device – are introduced in SFA system, if required, and they are connected in series to the other components of the manifold. For heating, usually thermostated baths or electrical wires wrapping the coils favoring the analytical reactions development are used. For separation, devices such as dialysers, liquid-liquid extractors, sorption or ion-exchange micro-columns, filters are used. These devices are placed before the reaction coils to remove potentially interfering species.

Debubbler – its aim is to remove the introduced air bubbles just before the liquid stream reaches the flow-cell of the detector. The removing is necessary in order to prevent the parasitic signals produced by the air bubbles upon reaching the detector. The debubbler is not normally necessary in the most recent designs as the signals from the detector are handled by a computer capable of discriminating between these undesirable signals and those actually corresponding to the reaction mixture.

Continuous detection system – any optical or electrochemical detector that can be equipped with a flow-cell is used as detection unit in SFA. The air bubbles are efficiently removed from the liquid stream if the waste tubing of the flow-cell is coupled to a channel of the peristaltic pump (Figure III.2.4). The flow rate of the liquid entering the flow-cell must be higher than the flow rate of the liquid drawn through the pump.

Data recording and treatment unit – which should be prepared to operate in continuous mode and be as simple as a typical Y–t recorder or a sophisticated as an advanced microprocessor/computer carrying out both operations or delivering directly the required results. The most modern SFA systems are fully operated by a high-performance computer.


Download 4.48 Mb.

Share with your friends:
1   2   3   4   5   6   7   8   9   ...   19




The database is protected by copyright ©ininet.org 2024
send message

    Main page