Automatic analytical methods for environmental monitoring and control



Download 4.48 Mb.
Page13/19
Date31.03.2018
Size4.48 Mb.
#44836
1   ...   9   10   11   12   13   14   15   16   ...   19

Figure III.2.31. A: absorbing solution: ADH, NAD+, phosphate buffer; P: peristaltic pump; D: fluorescence detector; V: injection valve. See explanation in the text.
The system was calibrated with the references:

QA= 0.23 mL min-1; QB= 0.15 mL min-1; gas flow rate=1 L min-1 and 70 % relative humidity. Calibration over the range 0 -112 mg mL-1. The dynamic range in not linear as usual in enzymatic reactions. The calculated deviation was 6.5 % (c = 44 mg mL-1, n = 5) and the required time per sample 2.3 min.


Determination of NO2 in air with on-line pre-concentration

In environmental air monitoring the level of analytes should be very low and the sample cannot be analyzed as is it in the environment, which obliges to include an on-line sample pre-treatment to obtain a continuous and automated procedure. The so-called chromato-membrane cell, CMC, has been used in different FIA systems for the on-line sample pre-concentration and separation (gas-liquid, liquid-liquid heterogeneous systems, etc.). It is a continuously working device in which the membrane performs a kind of “chromatographic separation”; when two different phases (polar and organic solvent or air) are forced through the membrane the transfer of analytes between tow phases occurs. This device is included in the FIA assembly.

The selected example to illustrate this is the determination of nitrogen dioxide in air so in field as in the laboratory. The analytical method is the “universal” for nitrite; first the nitrogen dioxide should be converted into nitrite. For this conversion is proposed the reported "chromato-membrane cell concentration-distribution device made of a PTFE block with micro- and macropores.

The steps of the flow determination are as follows (see Figure III.2.32.):



  1. The absorbing solution of triethanolamine is forced to the CMC at a flow-rate of 0.5 mL min-1 up to filling, and then the pump is stopped. A second pump is used to force the sample air (20 L) into the cell at 7 mL min-1. The NO2 is transferred from the air to the absorbing reagent and transformed into nitrite ions.

  2. Turning the valve 2 the resulting solution is inserted into the FIA system and merges with the mixture of sulfanilamide and NED (Saltzman reagent) developing the corresponding color to be monitored in the flow-cell of the spectrophotometer.

  3. By starting again in the reverse direction the free NO2 air enters into CMC cell to by renewed and start again the cycle.

The portable set-up is a micro FIA system with PTFE tubing 0.25 mm internal diameter; which is closely tight into a box (box size 16 x 16 x 32 cm) and the detector is of the LED type (light emitting diode) with a power source (12 V battery) (for laboratory determinations the FIA manifold was formed with PTFE tubing 0.5 mm internal diameter). The detection limit is 0.9 g L-1 and the complete cycle is about 5 min.
F
igure III.2.32.
Flow system with the chromato-membrane. RS, reagent solution (sulfanilamide and NED mixture); AS, absorbing solution; SL, standard sample (NaNO2); V, injection valve; P1, double-plunger pump (flow rate 0.05 and 0.25 mL min-1 for FIA or conventional FIA tubing, respectively); P2, peristaltic pump (0.5 mL min-1); P3, syringe type pump (7 mL min-1); DG, degassing unit; D, detector. CMC = CMC cell.
Ozone flow-analyzer

The authors do not clearly recommend a reagent; several were tested. They prepared an empirical set-up for a hanging droplet and a film-reaction and tested over 70 chemiluminescent reagents for ozone; the azine dyes phenosafranine, methylene blue and safranin O resulted in the higher light emission.

The main parts of the ozone analyzer are the detection zone:


  1. The hanging droplet is formed by a PTFE capillary 3 cm long and 200 nm internal diameter. The accurate and reproducible volume of solution is provided by the peristaltic pump; and the droplet is eliminated by the valve which presses briefly the pumping flexible tubing.

  2. the film-reagent is a smooth glass surface, 80 x 6 mm, provided with a triangular distributing device to avoid irregularities of the liquid film over the whole surface. Analytical applications are only studied in the film-reactor mode.

  3. The reactor consists of a 35 x 200 mm chamber build in black PVC. The ozone flowing at 2 L min-1 reacts with the suitable reagent (according to the operator choice).

  4. The chemiluminescent reagent is pumped to the detector device placed at about 20 mm of the entrance window of the Photon Counting Module.

The flow system was as shown Figure III.2.33. With 1 mmol L-1 phenosafranine in ethanol the following analytical figures were obtained: application range, 5.2-330 g m-3; quantification limit, 2.1 g m-3 and time per sample 5 s. Comparing results with the UV method this detector present lesser sample consumption and is faster and gives minor results; which can be due to the interferences from aromatic compounds which are not eliminated in the UV detector.
Determination of atmospheric SO2

It is one of the relevant polluting compounds to be controlled to test the air quality and many efforts and procedures have been proposed for its determination in air analyses. Most of these methods lack the required selectivity and suitable sensitivity. A pre-concentration step is the usual sample pre-treatment, which in the next selected example is performed with the aid of a gas permeation device. These devices have been extensively proposed in FIA procedures.

A spectrophotometric monitoring assembly provided with on-line sampling and pre-concentration and able to be field applied is reported.

A micro-pore hydrophobic membrane made of poly(vinyldene) difluoride, sandwiched between two perspex blocks serves for pre-concentration. Air circulates at one side of the membrane at a flow-rate of 0.9 L min-1; at the other side, the absorbing solution at 0.8 mL min-1 containing 5 10-4 mol L-1 5,5´-dithiobis(2,2´-dinitrobenzoic acid) (DTNB) in 0.025 mol L-1 phosphate buffer. For a 5 – 8 minutes interval the flow is stopped to allow the pre-concentration process.



The resulting mixture absorbed SO2 plus reagent is inserted by the injection valve into the carrier stream formed by the same reagent DNTB at 0.7 mL min-1 and is monitored in the detector flow-cell at 410 nm. A miniaturized optical fibber spectrophotometer is used (Figure III.2.34).





Figure III.2.33. Flow assembly for ozone determination. RC: Reaction chamber; D: photo counting module; V: valve; P: peristaltic pump

F
igure III.2.34.
FIA assembly (top) and pre-concentration device (lower) for SO2 determination
The main interferences of the procedure are the hydrogen sulfide and hydrogen cyanide.

The application range and detection limits rely on the pre-concentration time and the reference signal; e. g., for 5 min and reference the carrier DNTB solution, 0-4 mg m-3 and 50 g m-3, respectively. However, for 8 min of pre-concentration and reference output pure air, the analytical figures of merit are 0-3.2 mg m-3 and 35 g m-3. Sample throughput clearly linked to the pre-concentration interval, are 8.5 and 6 h-1 respectively.


III.2.3.6. Soil Pollutants
Determination of pesticides by a multi-commutation method with photo-chemiluminescence.

The pollution of underground waters by the extensive use of pesticides is a new anthropogenic problem affecting the health of the population. Pesticides are in the environment as a consequence of being used in agricultural works but also as being used in other activities as domestic use, airports, golf fields, roadsides, etc.

The massive use of pesticides (last two decades) is a new challenge in water treatment plants. Herbicides are the most used pesticides. The manufacturer’s numerical figures give an idea of the size of the problem. In Europe, more than 500 tons/year are fabricated of about 40 pesticides; the manufactured amount of pesticides in USA in 1993 is represented in the Table III.2.3.
Table III.2.3. Amount of pesticides produced in USA in 1993.


Pesticide

Tones

Pesticide

Tones

Atrazine

31,500-33,750

Chlorpyrifos

4,500-6,750

Metalachlor

27,000-29,250

Chlorothalonil

4,500-6,750

Alachlor

20,250-22,500

Propanil

3,150-5,400

Methyl bromide

13,500-15,750

Dicamba

2,700-4,500

Cyanazine

13,500-15,750

Terbufos

22,250-3,600

Dichloropropene

13,500-15,750

Bentazone

1,800-3,150

2,4-D

11,250-13,500

Mancozeb

1,800-3,150

Metam Sodium

11,250-13,500

Parathion

1,800-3,150

Trifluralin

9,000-11,250

Simazine

1,350-2,700

Glyphosate

6,750-9,000

Butylate

1,350-2,700

A pesticide is formulated with additives and co-adjuvant to facilitate its actuation. They are fumigated from air and its concentration in the environment is continuously changing due to dispersion, volatilization, chemical and biological degradation and lixiviation. How these processes are performed is due to the physic-chemical characteristics of each pesticide, but also from the characteristics of waters, soil and atmosphere.

There are many flow methods for pesticide determination. The selected example is based on a new strategy by means of the multi-commutation methodology; with the aid of the chemiluminescence detection and photo-degradation to form the detectable compound. The aqueous solution of the pesticide is irradiated with an UV lamp in the suitable medium. The formed photo-fragments merge with the oxidant, potassium permanganate in a sulfuric medium before the chemiluminescence-based detection of the resulting photoproducts. The use of solenoid valves results in substantial reagent savings and constitutes a further extension of clean chemistry procedures.

T
he pesticide asulam (methyl-4-aminobenzenesulphonyl carbamate), presents a broad spectrum of biological activity. It is used as an insecticide, herbicide and fungicide; and, most often, it is used as a post-emergency herbicide for controlling deciduous and perennial grasses. Asulam acts by stopping cell division and growth of plant tissues. It remains in soil for more than one season. However, it exhibits a high mobility by virtue of the high water solubility of its sodium salt and is therefore a potential water pollutant.



Monitoring photo-degradation processes has recently proved an effective method for the in situ control of environmental pollutants. The potential of sunlight-based photocatalytic decontamination has lately aroused much interest.

T
he flow manifold depicted in Figure III.2.35 comprised three solenoid valves each of which acted as a stand-alone commutator with only two positions. Valve operation was characterized in terms of N*(t1,t2), where t1 and t2 are the intervals during which the valve was ON and OFF, respectively, and N was the number of times of the ON/OFF cycle. The peristaltic pump was placed to aspirate the sample and reagents into the flow-cell behind the reactor; this difference from the usual location of the propulsion system in the FIA-manifolds, was intended to prevent the flow from stopping immediately upon insertion as the valve was actuated.

Figure III.2.35. Flow assembly optimized for pesticide determination.

Q2: photodegradation medium (glycine buffer at pH 8.3); Q1: Aqueous solution of asulam; Q4: Carrier (water); Q3: Oxidant (K3Fe(CN)6 0.1 molL-1 in NaOH 1 molL-1 or KMnO4 10-4 molL-1 in H2SO4 1.2 molL-1). Flow-rate: 9 and 10 mLmin-1 for Fe(CN)63- and MnO4- system, respectively. P: peristaltic pump; PMT: photomultiplier tube; V: Solenoid valve; FC: spiral flow cell; PR: photoreactor consisted of a 173 cm length and 0.8 mm i.d. PTFE tubing helically coiled around a 15 W low-pressure mercury lamp (Sylvania) for germicidal use.


The optimum insertion profile for each of the two tested oxidant systems is depicted in Figure III.2.36. With the potassium permanganate, an overall of 20 alternate micro-insertions of pesticide and photo­-degradation medium were performed. During each micro-insertion, valve V1 was kept ON for 0.3 s to aspirate asulam and OFF for 0.1 s to aspirate the buffer used as photo­-degradation medium. Valve V2 was kept on to have the peristaltic pump aspirate asulam and the buffered medium throughout the duration of the process (8 s). This loading time allowed the inner walls of the photo-reactor to be efficiently flushed in order to avoid sample carry-over. During the next 90 s, the sample/medium mixture was stopped in the photo-rector for the UV-irradiation. Next, valve V3 was switched ON for 17 s to allow t
he oxidant to flow and V2 was used to alternately micro-insert photo-­degraded pesticide (0.7 s segments) and the oxidant (0.2 s segments) in 17 ON/OFF cycles.

Figure III.2.36. Optimized insertion profile for obtaining a typical transient analytical signal for MnO4-/H2SO4 systems.
Two oxidant systems were studied: potassium permanganate and ferrycianide. With the selected MnO4/H2SO4 system, the response was linear up to an asulam concentration of 5 ppm; the relative standard deviation for the slope of five curves recorded for freshly made solutions on different days was 3.8%. This is a “clean” chemical procedure as the reagent uptake is very low in both cases (only 716 μL for potassium permanganate).

The MnO4/H2SO4 system proved more selective. Especially strong was the interference of calcium with the Fe(CN)63-/NaOH system and the photo-degradation of nitrate to nitrite which reacted with the permanganate. Copper gave a strong chemiluminescent signal, even at low concentrations. This may require the prior removal of copper from some types of sample with a solid-phase reactor filled with an appropriate ion-exchange resin. The limit of detection was 40 g L–1 asulam; the equation relating the two was I = 110.42 [mg L–1] – 214.84 (r2 = 0.9926). The overall analysis time was 120 s.


REFERENCES (ordered as in the text)


  1. Field-portable flow-injection analysers for monitoring of air and water pollution. P. W. Alexander, L. T. Di Benedetto, T. Dimitrakopoulos, D. B. Hibbert, J. C. Ngila, M. Sequeira and D. Shiels, Talanta, 1996, 43(6), 915 – 925.

  2. B. Karlberg, B. Pacey, Flow Injection Analysis, a practical guide, Elsevier, New York, 1989.

  3. Miniature flow injection analyser for laboratory, shipboard and in situ monitoring of nitrate in estuarine and coastal waters. P. C. F. C. Gardolinski, A. R. J. David and P. J. Worsfold, Talanta, 2002, 58(6), 1015 – 1027.

  4. A reverse-flow injection analysis method for the determination of dissolved oxygen in fresh and marine waters. S. Muangkaew, I. D. MaKelvie, M. R. Grace, M. Rayanakorn, K. Grudpan, J. Jakmunce and D. Nacapricha, Talanta, 2002, 58(6), 1285 – 1291.

  5. Flow-injection system with enzyme reactor for differential amperometric determination of hydrogen peroxide in rain water, R. Camargo Matos, J.J. Pedrotti and L. Angnes, Anal. Chim. Acta, 2001, 441(1), 73-79.

  6. Amperometric flow-injection technique for determination of hydrogen peroxide and sulphur(IV) in atmospheric liquid water, I.G.R. Gutz and D. Klockow, Fresenius' Z. Anal. Chem. 1989, 335(8), 919-923.

  7. Application of flowing stream techniques and related compounds to water analysis. Part I. Ionic species: dissolved inorganic carbon, nutrients, M. Miró, J. M. Estela and V. Cerdá, Talanta, 2003, 60(5), 867 – 886

  8. An enzymic-fluorimetric method for monitoring of ethanol in ambient air. M. Schilling, G. Voigt, T. Tavares and D. Klockow, Fresenius' J. Anal. Chem., 1999, 364(1-2), 100-105.

  9. Absorption, concentration and determination of trace amounts of air pollutants by flow injection method coupled with a chromatomembrane cell system: application to nitrogen dioxide determination. Y. L. Wei, M. Oshima, J. Simon, L.N. Moskvin and S. Motomizu, Talanta, 2002, 58(6), 1343-1355.

  10. Determination of ozone in ambient air with a chemiluminescence reagent film detector. C. Eipel, P. Jeroschewski and I. Steinke, Anal. Chim. Acta, 2003, 491(2), 145-153.

  11. Flow injection determination of gaseous sulfur dioxide with gas permeation denuder-based online sampling and preconcentration. Z.X. Guo, Y.Z. Li, X.X. Zhang, W.B. Chang and Y.X. Ci, Anal. Bioanal. Chem., 2002, 374(6), 1141-1146.

  12. A new flow-multicommutation method for the photo- chemiluminometric determination of the carbamate pesticide asulam. A. Chivulescu, M. Catalá-Icardo, J. V. García-Mateo and J. Martínez-Calatayud, Anal. Chim.Acta 2004, in print.




Download 4.48 Mb.

Share with your friends:
1   ...   9   10   11   12   13   14   15   16   ...   19




The database is protected by copyright ©ininet.org 2024
send message

    Main page